Barzani, Hemn A. H.Ali, Hoshyar SaadiYunusoglu, OrucYardim, Yavuz2025-05-102025-05-1020240925-96351879-006210.1016/j.diamond.2024.1109322-s2.0-85186494038https://doi.org/10.1016/j.diamond.2024.110932https://hdl.handle.net/20.500.14720/10911This investigation aims to describe the voltammetric analysis of ivacaftor (IVA) by utilizing the boron -doped diamond (BDD) electrode to add a cationic surfactant. Cyclic voltammetry via Britton-Robinson (BR, 0.04 mol L-1, pH 2.0) buffer was used to perform determinations of the IVA demonstrating irreversible behaviors, adsorption-controlled and well-defined (+1.04 V, PA1) and an ill-defined (+1.42 V, PA2) oxidation peaks (vs. Ag/ AgCl). The findings revealed that the oxidation peaks of IVA are pH-dependent (ranging from 2.0 to 5.0). The use of cetyltrimethylammonium bromide (CTAB, cationic surfactant) in the chosen supporting electrolyte enormously raised the oxidation peak currents of IVA. For the measurement of IVA in a 0.04 mol L-1 BR buffer solution with a pH of 2.0, the linear relationship was discovered to exist under the conditions of the experimental optimal parameters involving 2 x 10-4 mol L-1 CTAB at +1.11 V (vs. Ag/AgCl) (after an accumulation of 60 s at the open-circuit condition). The linear concentration was discovered using 0.25 to 10.0 mu g mL-1 (6.4 x 10-7-2.5 x 10-5 mol L-1) and the limit of detection 0.073 mu g mL-1 (1.9 x 10-7 mol L-1). The devised methodology was effectively employed to determine IVA in pharmaceutical formulation. To the best of our understanding, it represents the first electroanalytical method for detecting IVA by voltammetry.eninfo:eu-repo/semantics/closedAccessIvacaftorVoltammetric AnalysisCationic SurfactantPharmaceutical FormulationBoron -Doped Diamond ElectrodeSensing Ivacaftor Accomplished Using the Square-Wave Voltammetric Technique With the Assistance of a Cationic Surfactant on a Boron-Doped Diamond ElectrodeArticle144Q2Q2WOS:001198241800001