Dilmac, YusufGuler, Muhammet2025-05-102025-05-1020201572-66571873-256910.1016/j.jelechem.2020.1140912-s2.0-85082059158https://doi.org/10.1016/j.jelechem.2020.114091https://hdl.handle.net/20.500.14720/6042In this article, Au nanoparticles (AuNPs) were deposited on carboxylated graphene oxide (GO-COOH) for the electrochemical oxidation and enzyme-free voltammetric and amperometric determination of glucose. The GO-COOAu modified glassy carbon electrode (GCE) exhibited good sensitivity and stability toward glucose electro-oxidation in alkaline media. The surface morphology of the as-prepared nanocomposite was investigated by field emission scanning electron microscopy (FESEM) and X-ray diffraction (XRD). The electrochemical characterization of theworking electrodes was evaluated using cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS), and amperometry. The GO-COOAu/GCE sensor provided excellent electro-catalytic performance for glucose oxidation with a linear range from 0.02 to 4.48 mM (R-2 = 0.9919) at +0.35 V, a low limit of detection (LOD) of 6 mu M, and a high sensitivity of 20.218 mu A mM(-1) cm(-2). The GO-COOAu nanocompositemodified working electrode with high sensitivity, repeatability, reproducibility, and quantification of glucose in human serumsamples made it a promising material for the development of non-enzymatic sensors.eninfo:eu-repo/semantics/closedAccessCarboxylated Graphene OxideAu NanoparticlesGlucose SensorFabrication of Non-Enzymatic Glucose Sensor Dependent Upon Au Nanoparticles Deposited on Carboxylated Graphene OxideArticle864Q1Q2WOS:000528254900006