YYÜ GCRIS Basic veritabanının içerik oluşturulması ve kurulumu Research Ecosystems (https://www.researchecosystems.com) tarafından devam etmektedir. Bu süreçte gördüğünüz verilerde eksikler olabilir.
 

Fabrication of Non-Enzymatic Glucose Sensor Dependent Upon Au Nanoparticles Deposited on Carboxylated Graphene Oxide

No Thumbnail Available

Date

2020

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier Science Sa

Abstract

In this article, Au nanoparticles (AuNPs) were deposited on carboxylated graphene oxide (GO-COOH) for the electrochemical oxidation and enzyme-free voltammetric and amperometric determination of glucose. The GO-COOAu modified glassy carbon electrode (GCE) exhibited good sensitivity and stability toward glucose electro-oxidation in alkaline media. The surface morphology of the as-prepared nanocomposite was investigated by field emission scanning electron microscopy (FESEM) and X-ray diffraction (XRD). The electrochemical characterization of theworking electrodes was evaluated using cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS), and amperometry. The GO-COOAu/GCE sensor provided excellent electro-catalytic performance for glucose oxidation with a linear range from 0.02 to 4.48 mM (R-2 = 0.9919) at +0.35 V, a low limit of detection (LOD) of 6 mu M, and a high sensitivity of 20.218 mu A mM(-1) cm(-2). The GO-COOAu nanocompositemodified working electrode with high sensitivity, repeatability, reproducibility, and quantification of glucose in human serumsamples made it a promising material for the development of non-enzymatic sensors.

Description

Keywords

Carboxylated Graphene Oxide, Au Nanoparticles, Glucose Sensor

Turkish CoHE Thesis Center URL

WoS Q

Q1

Scopus Q

Q2

Source

Volume

864

Issue

Start Page

End Page