Effect of Thermoplastic Mask Hole Diameter on Dose Distribution in the Build-Up Region of Medical Linac
No Thumbnail Available
Date
2024
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
Abstract
This study aims to investigate the effect of variations in the hole diameter of unstretched thermoplastic masks offered by a single company on the dose distribution in the build-up region. 2, 2.5, 3 and 4 mm hole diameters masks with the same properties were irradiated with 6 MV X-rays and dose distribution was analyzed according to the open field. Dose measurements were obtained with a parallel plane ionization chamber. All irradiations were repeated for 5 x 5, 8 x 8, 10 x 10 and 15 x 15 cm 2 beam fields. It was observed that the mask had a bolus effect in all measurements. It was observed that this bolus effect varied especially depending on the hole diameter in the mask. In a 10 x 10 cm 2 field, the d max depth without mask was 15 mm, while the d max for masks with 2, 2.5, 3 and 4 mm hole diameters was 11, 12, 13 and 13 mm, respectively. It was observed that the mask hole diameter affected the d max depth. Thermoplastic masks with large hole diameters increase the surface dose less than those with small hole diameters.
Description
Demir, Hikmettin/0000-0002-1171-4821; Gul, Osman Vefa/0000-0002-6773-3132
Keywords
Radiotherapy, Thermoplastic Mask, Percentage Depth Dose, Surface Dose
Turkish CoHE Thesis Center URL
WoS Q
Q3
Scopus Q
N/A
Source
Volume
17
Issue
2