Effect of Thermoplastic Mask Hole Diameter on Dose Distribution in the Build-Up Region of Medical Linac

dc.authorid Demir, Hikmettin/0000-0002-1171-4821
dc.authorid Gul, Osman Vefa/0000-0002-6773-3132
dc.authorwosid Demir, Hikmettin/Hga-7385-2022
dc.authorwosid Gul, Osman Vefa/Aaf-2848-2021
dc.contributor.author Demir, Hikmettin
dc.contributor.author Gul, Osman Vefa
dc.date.accessioned 2025-05-10T17:23:55Z
dc.date.available 2025-05-10T17:23:55Z
dc.date.issued 2024
dc.department T.C. Van Yüzüncü Yıl Üniversitesi en_US
dc.department-temp [Demir, Hikmettin] Van Yuzuncu Yil Univ, Fac Med, Dept Radiat Oncol, Van, Turkiye; [Gul, Osman Vefa] Selcuk Univ, Fac Med, Dept Radiat Oncol, Konya, Turkiye en_US
dc.description Demir, Hikmettin/0000-0002-1171-4821; Gul, Osman Vefa/0000-0002-6773-3132 en_US
dc.description.abstract This study aims to investigate the effect of variations in the hole diameter of unstretched thermoplastic masks offered by a single company on the dose distribution in the build-up region. 2, 2.5, 3 and 4 mm hole diameters masks with the same properties were irradiated with 6 MV X-rays and dose distribution was analyzed according to the open field. Dose measurements were obtained with a parallel plane ionization chamber. All irradiations were repeated for 5 x 5, 8 x 8, 10 x 10 and 15 x 15 cm 2 beam fields. It was observed that the mask had a bolus effect in all measurements. It was observed that this bolus effect varied especially depending on the hole diameter in the mask. In a 10 x 10 cm 2 field, the d max depth without mask was 15 mm, while the d max for masks with 2, 2.5, 3 and 4 mm hole diameters was 11, 12, 13 and 13 mm, respectively. It was observed that the mask hole diameter affected the d max depth. Thermoplastic masks with large hole diameters increase the surface dose less than those with small hole diameters. en_US
dc.description.woscitationindex Science Citation Index Expanded
dc.identifier.doi 10.1016/j.jrras.2024.100884
dc.identifier.issn 1687-8507
dc.identifier.issue 2 en_US
dc.identifier.scopusquality N/A
dc.identifier.uri https://doi.org/10.1016/j.jrras.2024.100884
dc.identifier.uri https://hdl.handle.net/20.500.14720/11028
dc.identifier.volume 17 en_US
dc.identifier.wos WOS:001217209900001
dc.identifier.wosquality Q3
dc.language.iso en en_US
dc.publisher Elsevier en_US
dc.relation.publicationcategory Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/openAccess en_US
dc.subject Radiotherapy en_US
dc.subject Thermoplastic Mask en_US
dc.subject Percentage Depth Dose en_US
dc.subject Surface Dose en_US
dc.title Effect of Thermoplastic Mask Hole Diameter on Dose Distribution in the Build-Up Region of Medical Linac en_US