dc.description.abstract Treatment planning systems (TPS) have an important role in the implementation of radiotherapy treatment. Quality control of TPS is extremely important in terms of treatment reliability. In this study, it was aimed to make quality control of TPS by comparing the data obtained from the PROWESS Panther TPS with the results of the Monte Carlo simulation for small fields. In this study, TPS used in the radiation oncology department was examined. For MC simulation, BEAMnrc and DOSXYZnrc codes were used in modeling and dose measurement of linear accelerator device. Gamma analysis method was used to compare measurement results. When we examine the gamma analysis for PDDs in the study, the gamma values obtained in the 3x3 cm2, 4x4 cm2 and 5x5 cm2 field sizes for the 3% and 3mm criterion provide this criterion for each measurement point. The fact that each dose point in the measurement field is less than 0.1 gamma value indicates that the dose profiles obtained in PROWESS Panther TPS are in good agreement with the MC results. The doses calculated by PROWESS Panther TPS were found within the reference limits. It can be concluded that the dose calculated by PROWESS Panther TPS in homogeneous environments is within acceptable limits for use in radiotherapy. Studies should also be carried out in non-homogeneous environments such as lung, bone and soft tissue in order to obtain more accurate results about PROWESS Panther TPS. © 2022, Yuzuncu Yil Universitesi Tip Fakultesi. All rights reserved. en_US
dc.identifier.doi 10.5505/EJM.2022.88785
dc.identifier.endpage 56 en_US
dc.identifier.issn 1301-0883
dc.identifier.issue 1 en_US
dc.identifier.scopus 2-s2.0-85125854433
dc.identifier.scopusquality Q4
dc.identifier.startpage 53 en_US
dc.identifier.trdizinid 1130690
dc.identifier.uri https://doi.org/10.5505/EJM.2022.88785
dc.identifier.uri https://hdl.handle.net/20.500.14720/3013
dc.identifier.volume 27 en_US
dc.identifier.wosquality N/A
dc.institutionauthor Tuğrul, T.
dc.language.iso en en_US
dc.publisher Yuzuncu Yil Universitesi Tip Fakultesi en_US
dc.relation.ispartof Eastern Journal of Medicine en_US
dc.relation.publicationcategory Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/openAccess en_US
dc.subject Monte Carlo en_US
dc.subject Radiation en_US
dc.subject Radiation Oncology en_US
dc.subject Simulation Prowess Panther en_US
dc.subject Treatment Planning Systems en_US
dc.title Examination of Dosimetric Verification of Prowess Panther Treatment Planning System Using Monte Carlo Method for Small Fields en_US
dc.type Article en_US
dspace.entity.type Publication

Files