Dual Responsive Disposable Electrode for the Enumeration Ofescherichia Coliin Whole Blood

No Thumbnail Available

Date

2020

Journal Title

Journal ISSN

Volume Title

Publisher

Wiley-v C H verlag Gmbh

Abstract

In this report, we have designed a dual responsive disposable electrode for the enumeration ofEscherichia coli K12 (E. coli K12). The immunomagnetic separation strategy provided easily bacterial detection in the whole blood. Metal-organic frameworks magnetic (MOFs) modified, and citrate capped gold nanoparticles were used as capture probe and spectro-electrochemical labels, respectively. The cyclic voltammetry (CV) was employed to quantifyE. coliwith a ranging from 10(1)to 10(7) cfu/mL with a LOD of 1 cfu/mL. The Surface enhanced Raman spectroscopy (SERS) measurements were also performed with using disposable electrode.

Description

Tamer, Ugur/0000-0001-9989-6123; Ilhan, Hasan/0000-0002-4475-1629; Panhwar, Sallahuddin/0000-0001-9938-6369

Keywords

Disposable Electrode, Spectro-Electrochemical Labels, Cyclic Voltammetry, Pathogens Bacteria, Blood

Turkish CoHE Thesis Center URL

WoS Q

Q2

Scopus Q

Q2

Source

Volume

32

Issue

10

Start Page

2244

End Page

2252
Google Scholar Logo
Google Scholar™