Disposable Concentric Needle Electromyography Electrodes Artifact Discharge Mimicking Positive Sharp Waves

No Thumbnail Available

Date

2003

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

Objectives: To determine a new artifact discharge produced from the disposable concentric needle (DCN) electromyography (EMG) electrodes. Methods: We have recorded the activity obtained after the first insertion of 41 DCN (37 mm) and 36 DCN (50 mm) in several muscles during the rest. The number of the patients was 77 (26 male, 51 female). Results: We observed an artifact discharge (AD) resembling positive sharp waves (PSWs) in 31 patients of 77 insertions. The AD occurred with 18 of 41 insertions of DCN 37 mm and 13 of DCN 50 mm. The artifact resolved when the needle has been moved in the muscle or in the adipose subcutaneous tissue. Conclusion: This AD should be recognized when using DCN electrodes to avoid confusion with PSWs. Recognizing these artifacts during needle EMG is important to avoid false positive results.

Description

Keywords

Artifact Discharge, Disposable Concentric Needle, Electromyography

Turkish CoHE Thesis Center URL

WoS Q

N/A

Scopus Q

N/A

Source

Electromyography and Clinical Neurophysiology

Volume

43

Issue

1

Start Page

37

End Page

39