YYÜ GCRIS Basic veritabanının içerik oluşturulması ve kurulumu Research Ecosystems (https://www.researchecosystems.com) tarafından devam etmektedir. Bu süreçte gördüğünüz verilerde eksikler olabilir.
 

Evaluation of Effect of Different Insertion Speeds and Torques on Implant Placement Condition and Removal Torque in Polyurethane Dense D1 Bone Model

dc.authorscopusid 58654639800
dc.authorscopusid 55318687000
dc.authorwosid Orhan, Zeynep Dilan/Kfa-1989-2024
dc.contributor.author Orhan, Zeynep Dilan
dc.contributor.author Cigerim, Levent
dc.date.accessioned 2025-05-10T17:23:00Z
dc.date.available 2025-05-10T17:23:00Z
dc.date.issued 2024
dc.department T.C. Van Yüzüncü Yıl Üniversitesi en_US
dc.department-temp [Orhan, Zeynep Dilan; Cigerim, Levent] Van Yuzuncu Yil Univ, Fac Dent, Dept Oral & Maxillofacial Surg, TR-65090 Van, Turkiye en_US
dc.description.abstract The aim of this study was to evaluate the effect of two different insertion speeds at eight different insertion torque values ranging from 25 to 60 during implantation in a dense polyurethane (PU) D1 bone model on the placement condition and removal torque of dental implants. In this study, 50 pcf single-layer PU plates were used. In the study, a total of 320 implant sockets were divided into two groups, Group 1 (30 rpm) and Group 2 (50 rpm), in terms of insertion speed. Group 1 and Group 2 were divided into eight subgroups with 25, 30, 35, 40, 45, 50, 55 and 60 torques. There were 20 implant sockets in each subgroup. During the implantations, the implant placement condition and removal torque values were assessed. There was a statistically significant difference between the 30 and 50 rpm groups in terms of overall implant placement condition (p < 0.01). It was found that the removal torque values at 50 rpm were statistically significantly higher than those at 30 rpm (p < 0.01). This study showed that in dense D1 bone, the minimum parameters at which all implants could be placed at the bone level were 50 torque at 30 rpm and 40 torque at 50 rpm. en_US
dc.description.woscitationindex Science Citation Index Expanded
dc.identifier.doi 10.3390/polym16101361
dc.identifier.issn 2073-4360
dc.identifier.issue 10 en_US
dc.identifier.pmid 38794554
dc.identifier.scopus 2-s2.0-85194469767
dc.identifier.scopusquality Q1
dc.identifier.uri https://doi.org/10.3390/polym16101361
dc.identifier.uri https://hdl.handle.net/20.500.14720/10755
dc.identifier.volume 16 en_US
dc.identifier.wos WOS:001231429500001
dc.identifier.wosquality Q1
dc.language.iso en en_US
dc.publisher Mdpi en_US
dc.relation.publicationcategory Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/openAccess en_US
dc.subject Polyurethane Model en_US
dc.subject Polyurethane Foam en_US
dc.subject Polyurethane Plate en_US
dc.subject Dental Implant en_US
dc.subject Insertion Torque en_US
dc.subject Removal Torque en_US
dc.subject Implant Placement Condition en_US
dc.title Evaluation of Effect of Different Insertion Speeds and Torques on Implant Placement Condition and Removal Torque in Polyurethane Dense D1 Bone Model en_US
dc.type Article en_US

Files