YYÜ GCRIS Basic veritabanının içerik oluşturulması ve kurulumu Research Ecosystems (https://www.researchecosystems.com) tarafından devam etmektedir. Bu süreçte gördüğünüz verilerde eksikler olabilir.
 

Investigation of the Effect of Calibration Curves Obtained From Different Computed Tomography Devices on the Dose Distribution of Tomotherapy Plans

No Thumbnail Available

Date

2024

Journal Title

Journal ISSN

Volume Title

Publisher

Wolters Kluwer Medknow Publications

Abstract

Purpose: This study investigated the effect of Hounsfield units (HU)-relative electron density (RED) calibration curves obtained with devices from three different Computed Tomography (CT) manufacturers on dose distribution in Accuray Precision planning of patients with lung cancer. Methods: All CT data required for treatment planning system (TPS) were obtained using the Tomotherapy "cheese" phantom. HU RED calibration curves were created with images obtained from Siemens Somatom, GE Optima, and Toshiba Aquilion devices. The obtained calibration curve was extrapolated. CT images of lung cancer patients were acquired on a single device and treatment plans were created. The existing plans were recalculated using three calibration curves and the effect of the HU RED calibration curve on dose distribution was analyzed. Results: The results showed that different CTs did not produce significant dose differences in organ doses and PTV for Accuray TPS. Conclusions: Based on clinical judgment, images from different CT devices can be used in treatment planning.

Description

Demir, Hikmettin/0000-0002-1171-4821

Keywords

Calibration Curve, Computed Tomography, Hounsfield Unit, Tomotherapy

Turkish CoHE Thesis Center URL

WoS Q

N/A

Scopus Q

Q4

Source

Volume

49

Issue

4

Start Page

545

End Page

550