YYÜ GCRIS Basic veritabanının içerik oluşturulması ve kurulumu Research Ecosystems (https://www.researchecosystems.com) tarafından devam etmektedir. Bu süreçte gördüğünüz verilerde eksikler olabilir.
 

Investigation of the Effect of Calibration Curves Obtained From Different Computed Tomography Devices on the Dose Distribution of Tomotherapy Plans

dc.authorid Demir, Hikmettin/0000-0002-1171-4821
dc.authorscopusid 57193453129
dc.authorscopusid 59326935200
dc.authorscopusid 36810082100
dc.authorwosid Gul, Osman/Aaf-2848-2021
dc.authorwosid Demir, Hikmettin/Hga-7385-2022
dc.contributor.author Demir, Hikmettin
dc.contributor.author Gul, Osman Vefa
dc.contributor.author Kanyilmaz, Gul
dc.date.accessioned 2025-05-10T17:25:35Z
dc.date.available 2025-05-10T17:25:35Z
dc.date.issued 2024
dc.department T.C. Van Yüzüncü Yıl Üniversitesi en_US
dc.department-temp [Demir, Hikmettin] Yuzuncu Yil Univ, Dept Radiat Oncol, TR-65030 Van, Turkiye; [Gul, Osman Vefa] Selcuk Univ, Dept Radiat Oncol, Konya, Turkiye; [Kanyilmaz, Gul] Necmettin Erbakan Univ, Dept Radiat Oncol, Konya, Turkiye en_US
dc.description Demir, Hikmettin/0000-0002-1171-4821 en_US
dc.description.abstract Purpose: This study investigated the effect of Hounsfield units (HU)-relative electron density (RED) calibration curves obtained with devices from three different Computed Tomography (CT) manufacturers on dose distribution in Accuray Precision planning of patients with lung cancer. Methods: All CT data required for treatment planning system (TPS) were obtained using the Tomotherapy "cheese" phantom. HU RED calibration curves were created with images obtained from Siemens Somatom, GE Optima, and Toshiba Aquilion devices. The obtained calibration curve was extrapolated. CT images of lung cancer patients were acquired on a single device and treatment plans were created. The existing plans were recalculated using three calibration curves and the effect of the HU RED calibration curve on dose distribution was analyzed. Results: The results showed that different CTs did not produce significant dose differences in organ doses and PTV for Accuray TPS. Conclusions: Based on clinical judgment, images from different CT devices can be used in treatment planning. en_US
dc.description.woscitationindex Emerging Sources Citation Index
dc.identifier.doi 10.4103/jmp.jmp_129_24
dc.identifier.endpage 550 en_US
dc.identifier.issn 0971-6203
dc.identifier.issn 1998-3913
dc.identifier.issue 4 en_US
dc.identifier.pmid 39926133
dc.identifier.scopus 2-s2.0-85212915070
dc.identifier.scopusquality Q4
dc.identifier.startpage 545 en_US
dc.identifier.uri https://doi.org/10.4103/jmp.jmp_129_24
dc.identifier.uri https://hdl.handle.net/20.500.14720/11414
dc.identifier.volume 49 en_US
dc.identifier.wos WOS:001422004700007
dc.identifier.wosquality N/A
dc.language.iso en en_US
dc.publisher Wolters Kluwer Medknow Publications en_US
dc.relation.publicationcategory Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/openAccess en_US
dc.subject Calibration Curve en_US
dc.subject Computed Tomography en_US
dc.subject Hounsfield Unit en_US
dc.subject Tomotherapy en_US
dc.title Investigation of the Effect of Calibration Curves Obtained From Different Computed Tomography Devices on the Dose Distribution of Tomotherapy Plans en_US
dc.type Article en_US

Files